TE101 Tester
- Product name:TE101 Tester
- Category:SLT Testing Equipment
Product Features:
*Fully integrated and compact structure
* Suitable for single battery (DMFC&DFAFC)
* Meet the tests requirement of large area single battery
* Accurate gas and liquid flow control
* PC upper computer automatic operation
* Precise dedicated electronic load for fuel cell
*Multiple safety function
* Multifunctional interface
Application:
*Suitable for testing single cell and battery group of various fuel cells such as hydrogen oxygen, methanol and formic acid
Special service:
*Test system can be customized according to customer’s request, Increase the humidifying, AC impedance test, and Voltage inspection module, etc.
Product Features:
*Fully integrated and compact structure
* Suitable for single battery (DMFC&DFAFC)
* Meet the tests requirement of large area single battery
* Accurate gas and liquid flow control
* PC upper computer automatic operation
* Precise dedicated electronic load for fuel cell
*Multiple safety function
* Multifunctional interface
Application:
*Suitable for testing single cell and battery group of various fuel cells such as hydrogen oxygen, methanol and formic acid
Special service:
*Test system can be customized according to customer’s request, Increase the humidifying, AC impedance test, and Voltage inspection module, etc
Specification:
Parts System | Parameter | Remark |
Liquid Fuel Control System | ||
Liquid Flow Range | 0-1000ml/min | Software or Key setting |
Flow Control Accuracy | ±1% of FC | |
Negative Pole Gas Supply System | ||
Air Flow Range | 0-500SCCM | Software or Key setting |
Air Flow Test Accuracy | ±0.3%FS | |
Stack Thermal Management System | ||
Model | Positive and negative extreme plate heating control | Equipped with 4 pieces of 50W ceramic heating rods |
Extreme Plate Heating Power | 220V AC/200W | |
Temperature Control Range | Normal~99℃ | |
Temperature Control Accuracy | ±1℃ | Smart PID,Software or Key setting |
Cell Temperature Test Range | 0~300℃ | |
Electronic load | ||
Rated Power | 75W | |
Current Range | 0~15A | |
Voltage Range | 0~150V | |
Measure Accuracy | ±0.2%FS; ±0.2% SET POINT | |
Current Corresponding Time | Rise time<10us | High velocity response, Suitable For Fuel Cell Transient Analysis |
Safety Protection System | ||
Security Alarm | Safe upper limit for over-temperature and input voltage, Lower limit of protection for cell voltage | Brake application, Automatic load breaking, stop heating power |
Protection Method | Loop insurance, System emergency braking | |
Software Controls and Data Acquisition System | ||
Test Software For Fuel Cell Control System | Ware 3.0(LABVIEW) | Function of any - time high - speed, can generate associated graphs and data acquisition system as needed, TXT or Excel to save(Temperature, Flow, Voltage, Current, Test Time, Etc) |
Hydrogen Energy Category
- MEA material system
- Fuel Cell MEA System
- Fuel Cell Stacks System
- Fuel Cell Power Supply System
- Fuel Cell Teaching Instrument System
- Fuel Cell Special Equipment System
- Fuel Cell Vehicle System
Solar Wind Hybrid Category
- Solar&Wind Hybrid System
- Wind Turbine Generator System
- Smart Solar Lighting System
- Solar & Wind Hybrid Solution
New Smart Solar Category
- Solar Material Series
- DSSC Cell Series
- Perovskite solar cell Series
- Solar Cell Module Series
- Solar Equipment Series
Smart Lighting Category
- Induction Lighting System
- Grow Light systerm
- Road Lighting System
- Flood Lighing System
- Landscape Lighting System
- Indoors Lighting System
- Industrial Lighting System